					DC WITHSTAND VOLTAGE (7120,7122 ONLY)
				 | 
		
			| 
					Output Rating
				 | 
					6KVDC/5mA
				 | 
		
			| 
					Output Voltage, KVDC
				 | 
					0 - 6.00
				 | 
					0.01
				 | 
					± ( 2% of setting + 5V )
				 | 
		
			| 
					Output Ripple
				 | 
					<5% ( 6KV/5mA at Resistive Load )
				 | 
		
			| 
					SETTINGS
				 | 
		
			| 
					Hi-Limit DC Current, mA
				 | 
					0.02 - 5.00
				 | 
					0.01
				 | 
					± ( 2% of setting + 2counts )
				 | 
		
			| 
					Lo-Limit DC Current, mA
				 | 
					0.00 - 5.00
				 | 
					0.01
				 | 
					± ( 2% of setting + 2counts )
				 | 
		
			| 
					Ramp Time, second
				 | 
					0.1 - 999.9
				 | 
					0.1
				 | 
					± ( 0.1% + 0.05sec )
				 | 
		
			| 
					Dwell Time, second
				 | 
					0, 0.2 - 999.9
				 | 
					0.1
				 | 
		
			| 
					(0=continuous)
				 | 
		
			| 
					Arc Detection
				 | 
					0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
				 | 
		
			| 
					Discharge Time
				 | 
					< 200 msec
				 | 
		
			| 
					INSULATION RESISTANCE (7112,7122 ONLY)
				 | 
		
			| 
					Output Rating
				 | 
					1KVDC/1000MΩ
				 | 
		
			| 
					Output Voltage, KVDC
				 | 
					0.10 - 1.00
				 | 
					0.01
				 | 
					± ( 2% of setting + 3V )
				 | 
		
			| 
					SETTINGS
				 | 
		
			| 
					Hi-Limit Resistance, MΩ
				 | 
					0, 1 - 1000 (0=OFF)
				 | 
					1
				 | 
					± ( 3% of setting+2counts ) at > 500VDC
				 | 
		
			| 
					Lo-Limit Resistance, MΩ
				 | 
					1 - 1000
				 | 
					1
				 | 
					± ( 7% of setting +2counts ) at <500VDC
				 | 
		
			| 
					Delay Time, second
				 | 
					0, 0.5 - 999.9
				 | 
					0.1
				 | 
					± ( 0.1% + 0.05sec )
				 | 
		
			| 
					(0=continuous)
				 | 
		
			| 
					MEASUREMENT
				 | 
		
			| 
					AC/DC Voltage, KV
				 | 
					0 - 6.00
				 | 
					0.01
				 | 
					± ( 1.5% of reading + 1count )
				 | 
		
			| 
					DC Voltage, KV (IR only)
				 | 
					0.10 - 1.00
				 | 
					0.01
				 | 
					± ( 1.5% of reading + 3V )
				 | 
		
			| 
					AC Current, mA
				 | 
					0.10 - 12.00
				 | 
					0.01
				 | 
					± ( 2% of reading + 2counts )
				 | 
		
			| 
					DC Current, mA
				 | 
					0.02 - 5.00
				 | 
					0.01
				 | 
					± ( 2% of reading + 2counts )
				 | 
		
			| 
					Resistance, MΩ
				 | 
					1 - 1000
				 | 
					0.001
				 | 
					± ( 3% of reading+2counts ) at > 500VDC
				 | 
		
			| 
					(Auto Range)
				 | 
					0.01
				 | 
					± ( 7% of reading +2counts ) at <500VDC
				 | 
		
			| 
					 
				 | 
					0.1
				 | 
					 
				 | 
		
			| 
					GENERAL
				 | 
		
			| 
					Input Voltage AC
				 | 
					115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A Fast
				 | 
		
			| 
					PLC Remote Control
				 | 
					Input : Test , Reset
				 | 
		
			| 
					Output : Pass , Fail , Processing
				 | 
		
			| 
					Memory
				 | 
					5 Sets Store/Recall of each Memory, 4 Test Mode Selectable ( W, I, W ~ I, I ~ W Only 7112, 7122 )
				 | 
		
			| 
					Display
				 | 
					16×2 LCD with back light
				 | 
		
			| 
					Key Lock
				 | 
					To prevent unauthorized alteration of the test Parameters
				 | 
		
			| 
					Calibration
				 | 
					Build-in software and external calibrated meters
				 | 
		
			| 
					Environment
				 | 
					0-40℃, 20-80%RH
				 | 
		
			| 
					Dimension/Net Weight
				 | 
					280mm (W)×89mm(H)×370mm(D)/ 9Kg
				 | 
		
			| 
					STANDARD ACCESSORIES
				 | 
		
			| 
					Power Cord(10A)
				 | 
					×1
				 | 
		
			| 
					Fuses
				 | 
					×3 (Including a spare contained in the fuse holder)
				 | 
		
			| 
					Hipot Test Lead ,1.5m(1101)
				 | 
					×1
				 | 
		
			| 
					Hipot Return Lead ,1.8m(1102)
				 | 
					×1
				 | 
		
			| 
					*product specifications are subject to change without notice.
				 |