						SPECIFICATIONS 
					 | 
			
				| 
						MODEL
					 | 
						7151
					 | 
						7152
					 | 
						7160
					 | 
						7162
					 | 
			
				| 
						AC WITHSTAND VOLTAGE
					 | 
			
				| 
						Output Rating
					 | 
						5KVAC/12mA
					 | 
			
				| 
						 
					 | 
						Range
					 | 
						Resolution
					 | 
						Accuracy
					 | 
			
				| 
						Output Voltage, KVAC
					 | 
						0-5.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 5V )
					 | 
			
				| 
						Output Frequency
					 | 
						50Hz/60Hz ±100ppm, User Selection
					 | 
			
				| 
						Output Waveform
					 | 
						Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
					 | 
			
				| 
						Output Regulation
					 | 
						± (1% of output + 5V), From no load to full load
					 | 
			
				| 
						SETTINGS
					 | 
			
				| 
						Hi-Limit AC Current, mA
					 | 
						0.10 - 12.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 2counts )
					 | 
			
				| 
						Lo-Limit AC Current, mA
					 | 
						0.00 - 12.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 2counts )
					 | 
			
				| 
						Ramp Time, second
					 | 
						0.1 - 999.9
					 | 
						0.1
					 | 
						± ( 0.1% + 0.05sec )
					 | 
			
				| 
						Dwell Time, second
					 | 
						0, 0.2 - 999.9
					 | 
						0.1
					 | 
			
				| 
						(0=continuous)
					 | 
			
				| 
						Arc Detection
					 | 
						0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
					 | 
			
				| 
						DC WITHSTAND VOLTAGE (7120,7122 ONLY)
					 | 
			
				| 
						Output Rating
					 | 
						6KVDC/5mA
					 | 
			
				| 
						Output Voltage, KVDC
					 | 
						0 - 6.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 5V )
					 | 
			
				| 
						Output Ripple
					 | 
						<5% ( 6KV/5mA at Resistive Load )
					 | 
			
				| 
						SETTINGS
					 | 
			
				| 
						Hi-Limit DC Current, mA
					 | 
						0.02 - 5.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 2counts )
					 | 
			
				| 
						Lo-Limit DC Current, mA
					 | 
						0.00 - 5.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 2counts )
					 | 
			
				| 
						Ramp Time, second
					 | 
						0.1 - 999.9
					 | 
						0.1
					 | 
						± ( 0.1% + 0.05sec )
					 | 
			
				| 
						Dwell Time, second
					 | 
						0, 0.2 - 999.9
					 | 
						0.1
					 | 
			
				| 
						(0=continuous)
					 | 
			
				| 
						Arc Detection
					 | 
						0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
					 | 
			
				| 
						Discharge Time
					 | 
						< 200 msec
					 | 
			
				| 
						INSULATION RESISTANCE (7112,7122 ONLY)
					 | 
			
				| 
						Output Rating
					 | 
						1KVDC/1000MΩ
					 | 
			
				| 
						Output Voltage, KVDC
					 | 
						0.10 - 1.00
					 | 
						0.01
					 | 
						± ( 2% of setting + 3V )
					 | 
			
				| 
						SETTINGS
					 | 
			
				| 
						Hi-Limit Resistance, MΩ
					 | 
						0, 1 - 1000 (0=OFF)
					 | 
						1
					 | 
						± ( 3% of setting+2counts ) at > 500VDC
					 | 
			
				| 
						Lo-Limit Resistance, MΩ
					 | 
						1 - 1000
					 | 
						1
					 | 
						± ( 7% of setting +2counts ) at <500VDC
					 | 
			
				| 
						Delay Time, second
					 | 
						0, 0.5 - 999.9
					 | 
						0.1
					 | 
						± ( 0.1% + 0.05sec )
					 | 
			
				| 
						(0=continuous)
					 | 
			
				| 
						MEASUREMENT
					 | 
			
				| 
						AC/DC Voltage, KV
					 | 
						0 - 6.00
					 | 
						0.01
					 | 
						± ( 1.5% of reading + 1count )
					 | 
			
				| 
						DC Voltage, KV (IR only)
					 | 
						0.10 - 1.00
					 | 
						0.01
					 | 
						± ( 1.5% of reading + 3V )
					 | 
			
				| 
						AC Current, mA
					 | 
						0.10 - 12.00
					 | 
						0.01
					 | 
						± ( 2% of reading + 2counts )
					 | 
			
				| 
						DC Current, mA
					 | 
						0.02 - 5.00
					 | 
						0.01
					 | 
						± ( 2% of reading + 2counts )
					 | 
			
				| 
						Resistance, MΩ
					 | 
						1 - 1000
					 | 
						0.001
					 | 
						± ( 3% of reading+2counts ) at > 500VDC
					 | 
			
				| 
						(Auto Range)
					 | 
						0.01
					 | 
						± ( 7% of reading +2counts ) at <500VDC
					 | 
			
				| 
						 
					 | 
						0.1
					 | 
						 
					 | 
			
				| 
						GENERAL
					 | 
			
				| 
						Input Voltage AC
					 | 
						115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A Fast
					 | 
			
				| 
						PLC Remote Control
					 | 
						Input : Test , Reset
					 | 
			
				| 
						Output : Pass , Fail , Processing
					 | 
			
				| 
						Memory
					 | 
						5 Sets Store/Recall of each Memory, 4 Test Mode Selectable ( W, I, W ~ I, I ~ W Only 7112, 7122 )
					 | 
			
				| 
						Display
					 | 
						16×2 LCD with back light
					 | 
			
				| 
						Key Lock
					 | 
						To prevent unauthorized alteration of the test Parameters
					 | 
			
				| 
						Calibration
					 | 
						Build-in software and external calibrated meters
					 | 
			
				| 
						Environment
					 | 
						0-40℃, 20-80%RH
					 | 
			
				| 
						Dimension/Net Weight
					 | 
						280mm (W)×89mm(H)×370mm(D)/ 9Kg
					 | 
			
				| 
						STANDARD ACCESSORIES
					 | 
			
				| 
						Power Cord(10A)
					 | 
						×1
					 | 
			
				| 
						Fuses
					 | 
						×3 (Including a spare contained in the fuse holder)
					 | 
			
				| 
						Hipot Test Lead ,1.5m(1101)
					 | 
						×1
					 | 
			
				| 
						Hipot Return Lead ,1.8m(1102)
					 | 
						×1
					 | 
			
				| 
						*product specifications are subject to change without notice.
					 |